Tips for EEG recording using Enobio or StarStim

In this page we provide some tips and tricks to work with our systems. Here are some useful resources you should also check:

- | StarStim tCS EEG recording (Jove)

- | A short introduction to NIC v1.2

- | Stimulation configuration with NIC v1.2


Plan your session well before starting

You should have a clear plan covering all the details in your recording session. Prepare to be patient. Setting up always takes a bit longer than expected, and when carrying out experiments ... the third time's the charm!


Where do you want to place the electrodes? For how long do you want to record? Do you have a good naming scheme for your EEG files? What data formats do you want to save? Do you need accelerometry data? Do you want a line filter applied to the data (if so 50 or 60 Hz?)?

Do you have an electrode referencing scheme? You should include in your montage a solution for data referencing. Data will be recorded referenced to the device electrical ground (the CMS electrode), but for data analysis you should always reference the data to an electrode or to the electrode average.


Place carefully the CMS and DRL electrodes

These two electrodes should go over the left or right mastoid (best to always stick to one side). First clean up the mastoid area where you are going to attach the DRL/CMS electrodes (using Sticktrode adhesive electrodes). You can use a paper napkin with some water or alcohol. Removing grease and dry skin will help you get good signal.

Stick the DRL/CMS electrodes close to each other, with the DRL on the bottom. These two Sticktrode electrodes should be close to each other but not touch. The CMS electrode should be on top and squarely on top of the mastoid bone. This will avoid contamination from blood vessels (ECG like signals).

Avoid loose wires

Try to have a tidy setup with few loose wires. Loose wires are more prone to creating noise from movement.


About EEG signal quality measures in NIC

As of NIC v1.2, EEG signal quality measures for each electrode are provided in real time.

If all the channels are in yellow or red zone, you may have an issue with the CMS/DRL electrodes (they create the ground for measurements, so if they are misplaced all the signals will be affected).

If an electrode is misbehavig, check that it has a good contact and, if possible, add more gel (you can reach the site from another hole in the cap.

Tips for great Stimulation sessions using StarStim

Using Sponge electrodes

Sponge electrodes are easy to use. You can place them on the cap and wet them (about 2-4 cubic cm of saline solution will do) prior placing the cap on the subject's head.

If you have impedance issues, rewet them from a side hole using a syringe. Be careful not to wet the cap.

Using Ag/AgCl Pi electrodes

Pi electrodes are also easy to use. You can place them on the cap and gel them (fill the electrode hole space and then some more) prior placing the cap on the subject's head.

If you have impedance issues, regel them from the electrode hole or a side hole using a syringe.

The impedance test

You should carry out an impedance test before launching the stimulation. If impedance is too high for some electrode, rewet, regel and check cabling connections.

Impedance is check while stimulation is ongoing. If the impedance is too high, the stimulation session will self-abort.

Everything about electrodes for EEG and stimulation

Electrodes1.png

Please see the | Electrode User Manual for a description of our electrodes.

Caring for EEG electrodes

Electrodes2.png


Caring for stimulation electrodes

Electrodes3.png

It's important to rinse the electrodes with water in order to remove the salt that might remain after drying. It is also recommended to dry the metal part to avoid salt on the connectors.

In case there salt accumulating at the metallic part, it can be washed with some vinegar to remove it.

Electrode duration